AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 465

Error Code System Message

465

AMX-4, AMX-4+, Definium AMX 700, ERROR 465, Error 465
Question

What should I check for error 465 on the AMX-4 Series?

Short Answer

1) Check J2 plug from T2 on AMX1A4 panel and clean J3 on Filament kVp Board. 2) Filament kVp Board may be defective. 3) Batteries may be soft. Perform load test. BATTERY VOLTAGE MUST BE 113 V OR LESS TO DO LOAD TEST; DO NOT CHARGE BATTERIES BEFORE LOAD TEST. a) Connect oscilloscope from circuit breaker to ground; use Sample mode. b) Make exposures at 80, 100 and 125 kVp, about 10 mAs. c) Read battery voltage and average ripple at bottom. d) Battery voltage should drop about 13 VDC +/-10% at 80 kVp. e) About 16 VDC +/-10% at 100 kVp. f) About 17 VDC +/-10% at 125 kVp. 4) If batteries pass load test, unit may require recalibration of mAs, kVp, taps and filament current tables. 5) If batteries do not pass load test, replace batteries.

Possible Causes

1. Check J2 Plug From T2 On AMX1A4 Panel/Clean J3 On Filament kVp Board 2. Filament kVp Board may be defective 3. Batteries May Be Soft. Perform The Load Test. BATTERY VOLTAGE MUST BE 113 VOLTS OR LESS TO DO LOAD TEST! DO NOT CHARGE BATTERIES BEFORE LOAD TEST!! a. Connect Oscilloscope From Circuit Breaker To Ground - Use Sample Mode b. Make Exposures At 80, 100 kVp And 125 kVp, About 10 mAs c. Read Battery Voltage And Average The Ripple At The Bottom d. Battery Voltage Should Drop About 13 VDC +/-10% At 80 kVp e. Battery Voltage Should Drop About 16 VDC +/-10% At 100kVp f. Battery Voltage Should Drop About 17 VDC +/-10% At 125kVp 4. If Batteries Pass Load Test, Unit May Require Recalibration Of mAs, kVp, Taps And Filament Current Tables 5. If Batteries Do Not Pass The Load Test, replace batteries

Troubleshooting Steps

1. Check J2 Plug From T2 On AMX1A4 Panel/Clean J3 On Filament kVp Board 2. Filament kVp Board may be defective 3. Batteries May Be Soft. Perform The Load Test. BATTERY VOLTAGE MUST BE 113 VOLTS OR LESS TO DO LOAD TEST! DO NOT CHARGE BATTERIES BEFORE LOAD TEST!! a. Connect Oscilloscope From Circuit Breaker To Ground - Use Sample Mode b. Make Exposures At 80, 100 kVp And 125 kVp, About 10 mAs c. Read Battery Voltage And Average The Ripple At The Bottom d. Battery Voltage Should Drop About 13 VDC +/-10% At 80 kVp e. Battery Voltage Should Drop About 16 VDC +/-10% At 100kVp f. Battery Voltage Should Drop About 17 VDC +/-10% At 125kVp 4. If Batteries Pass Load Test, Unit May Require Recalibration Of mAs, kVp, Taps And Filament Current Tables 5. If Batteries Do Not Pass The Load Test, replace batteries

Parts Areas to Check

Filament kVp Board; Battery

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 465

Error Code System Message

ERROR 465 - Filament-current command feedback low during exposure interlock

ERROR 465, Filament-current command feedback low during exposure interlock, AMX-4 ERROR 465, AMX-4+ ERROR 465, Definium AMX 700 ERROR 465, GE AMX error ERROR 465
Question

What does ERROR 465 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 465: Filament-current command feedback low during exposure interlock.

Possible Causes

Filament-current command feedback was below expected during exposure interlock.

Troubleshooting Steps

Check for excessive battery voltage drop, filament feedback and generator control.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 466

Error Code System Message

ERROR 466 - Hardware backup timer expired

ERROR 466, Hardware backup timer expired, AMX-4 ERROR 466, AMX-4+ ERROR 466, Definium AMX 700 ERROR 466, GE AMX error ERROR 466
Question

What does ERROR 466 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 466: Hardware backup timer expired.

Possible Causes

Exposure hardware backup timing limit expired.

Troubleshooting Steps

Check backup timer status/oscillator and technique duration; review battery/timing conditions.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 466

Error Code System Message

466

AMX-4, AMX-4+, Definium AMX 700, ERROR 466, Error 466
Question

What should I check for error 466 on the AMX-4 Series?

Short Answer

1) During calibration, take unit out of calibration and reset CPU, then restart Generator Calibration (mAs, kVp, taps and filament current table). 2) Check spark gaps on HV Transformer Tank for 0.010 in. setting. 3) Check for miswired or open plug J2 from T2 on AMX1A4 panel to HV Transformer Tank mA+ & mA-. 4) Check for HV Transformer Tank spits. 5) Check for X-Ray Tube HV spits. 6) CPU Board may be defective. 7) On older units, insulate capacitor C5 from its retaining clamp (Service Note 3265).

Possible Causes

1. During Calibration, Take Unit Out Of Calibration And Reset CPU Then Restart Generator Calibration (mAs, kVp, Taps, And Filament Current Table) 2. Check Spark Gaps On HV Transformer Tank For A 0.010” Setting 3. Check for miswired Or Open Plug J2 From T2 On AMX1A4 Panel To HV Transformer Tank MA+ & MA- 4. Check for HV Transformer Tank Spits 5. Check for XRT HV Spits 6. CPU Board may be defective 7. Insulate (On Older Units) Capacitor C5 From its Retaining Clamp (Refer to Service Note 3265)

Troubleshooting Steps

1. During Calibration, Take Unit Out Of Calibration And Reset CPU Then Restart Generator Calibration (mAs, kVp, Taps, And Filament Current Table) 2. Check Spark Gaps On HV Transformer Tank For A 0.010” Setting 3. Check for miswired Or Open Plug J2 From T2 On AMX1A4 Panel To HV Transformer Tank MA+ & MA- 4. Check for HV Transformer Tank Spits 5. Check for XRT HV Spits 6. CPU Board may be defective 7. Insulate (On Older Units) Capacitor C5 From its Retaining Clamp (Refer to Service Note 3265)

Parts Areas to Check

CPU Board; HV Transformer Tank

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 467

Error Code System Message

467

AMX-4, AMX-4+, Definium AMX 700, ERROR 467, Error 467
Question

What should I check for error 467 on the AMX-4 Series?

Short Answer

1. Capacitor C5 Leaky AMX-4 Units (Does Not Apply To AMX-4+ Units) 2. Defective CPU Board 3. Bad battery Cell(s) 4. Defective Filament kVp Control Board 5. Unit recalibration is required (If In The “CHECKSUM-BYPASS” Mode) 6. HV Cable Open/Shorted Internally Or Poor Pin Spread On HV Candles

Possible Causes

1. Capacitor C5 Leaky AMX-4 Units (Does Not Apply To AMX-4+ Units) 2. Defective CPU Board 3. Bad battery Cell(s) 4. Defective Filament kVp Control Board 5. Unit recalibration is required (If In The “CHECKSUM-BYPASS” Mode) 6. HV Cable Open/Shorted Internally Or Poor Pin Spread On HV Candles

Troubleshooting Steps

1. Capacitor C5 Leaky AMX-4 Units (Does Not Apply To AMX-4+ Units) 2. Defective CPU Board 3. Bad battery Cell(s) 4. Defective Filament kVp Control Board 5. Unit recalibration is required (If In The “CHECKSUM-BYPASS” Mode) 6. HV Cable Open/Shorted Internally Or Poor Pin Spread On HV Candles

Parts Areas to Check

CPU Board; Battery

For Related Models

AMX-4, AMX-4+, Definium AMX 700