AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0x0100
Error Code System Message
Question
What does 0x0100 mean on a GE Brivo XR285amx?
Short Answer
0x0100: Unknown FW error.
Possible Causes
The Djinn Beetle generator reported a Heater 1 FW fault: Unknown FW error.
Troubleshooting Steps
Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0101
Error Code System Message
Question
What does 0x0101 mean on a GE Brivo XR285amx?
Short Answer
0x0101: Soft over current (measured current > estimated current).
Possible Causes
The Djinn Beetle generator reported a Heater 1 FW fault: Soft over current (measured current > estimated current).
Troubleshooting Steps
Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0102
Error Code System Message
Question
What does 0x0102 mean on a GE Brivo XR285amx?
Short Answer
0x0102: Soft open circuit (measured current < estimated current).
Possible Causes
The Djinn Beetle generator reported a Heater 1 FW fault: Soft open circuit (measured current < estimated current).
Troubleshooting Steps
Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0103
Error Code System Message
Question
What does 0x0103 mean on a GE Brivo XR285amx?
Short Answer
0x0103: Demand too high (target > max heater).
Possible Causes
The Djinn Beetle generator reported a Heater 1 FW fault: Demand too high (target > max heater).
Troubleshooting Steps
Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0103
Error Code System Message
Question
What does 0x0906 mean on a GE Brivo XR285amx?
Short Answer
c: Application not compatible with Boot.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: Application not compatible with Boot.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check