top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0100

Error Code System Message

​

0x0100, 0X0100, Brivo XR285amx 0x0100, Djinn Heater 1 error, Unknown FW error
Question

What does 0x0100 mean on a GE Brivo XR285amx?

Short Answer

0x0100: Unknown FW error.

Possible Causes

The Djinn Beetle generator reported a Heater 1 FW fault: Unknown FW error.

Troubleshooting Steps

Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0101

Error Code System Message

​

0x0101, 0X0101, Brivo XR285amx 0x0101, Djinn Heater 1 error, Soft over current (measured current > estimated current)
Question

What does 0x0101 mean on a GE Brivo XR285amx?

Short Answer

0x0101: Soft over current (measured current > estimated current).

Possible Causes

The Djinn Beetle generator reported a Heater 1 FW fault: Soft over current (measured current > estimated current).

Troubleshooting Steps

Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0102

Error Code System Message

​

0x0102, 0X0102, Brivo XR285amx 0x0102, Djinn Heater 1 error, Soft open circuit (measured current
Question

What does 0x0102 mean on a GE Brivo XR285amx?

Short Answer

0x0102: Soft open circuit (measured current < estimated current).

Possible Causes

The Djinn Beetle generator reported a Heater 1 FW fault: Soft open circuit (measured current < estimated current).

Troubleshooting Steps

Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0103

Error Code System Message

​

0x0103, 0X0103, Brivo XR285amx 0x0103, Djinn Heater 1 error, Demand too high (target > max heater)
Question

What does 0x0103 mean on a GE Brivo XR285amx?

Short Answer

0x0103: Demand too high (target > max heater).

Possible Causes

The Djinn Beetle generator reported a Heater 1 FW fault: Demand too high (target > max heater).

Troubleshooting Steps

Run Fil 1 Drv Clsd Loop, Fil 1 Xfmr Primary, and Fil 1 Drv Short Test as applicable. Check the J5/J6 filament power connection, cathode HV cable, filament impedance, and 48 V supply before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0103

Error Code System Message

​

0x0906, 0X0906, Brivo XR285amx 0x0906, Djinn Boot error, Application not compatible with Boot
Question

What does 0x0906 mean on a GE Brivo XR285amx?

Short Answer

c: Application not compatible with Boot.

Possible Causes

The Djinn Beetle generator reported a Boot SW fault: Application not compatible with Boot.

Troubleshooting Steps

Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page