AMX Troubleshooting & Technical Support
Results Count
Error Number
CAL TAP ERROR 2
Error Code System Message
CAL TAP ERROR 2 - High-kVp tap voltage too high
Question
What does CAL TAP ERROR 2 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL TAP ERROR 2: High-kVp tap voltage too high.
Possible Causes
Tube-voltage port exceeded the allowed high-kVp range.
Troubleshooting Steps
Repeat kVp calibration and inspect tap relay wiring.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CAL TAP ERROR 3
Error Code System Message
CAL TAP ERROR 3 - Low-kVp tap voltage too high
Question
What does CAL TAP ERROR 3 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL TAP ERROR 3: Low-kVp tap voltage too high.
Possible Causes
Tube-voltage port exceeded the allowed low-kVp range.
Troubleshooting Steps
Repeat kVp calibration and inspect generator/tap wiring.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CAL TAP ERROR 4
Error Code System Message
CAL TAP ERROR 4 - Tap combination could not be selected
Question
What does CAL TAP ERROR 4 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL TAP ERROR 4: Tap combination could not be selected.
Possible Causes
No acceptable tap combination could meet calibration requirements.
Troubleshooting Steps
Check generator/tap relay operation and repeat Tap Calibration.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CAL TAP ERROR 5
Error Code System Message
CAL TAP ERROR 5 - Filament-current feedback out of range
Question
What does CAL TAP ERROR 5 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL TAP ERROR 5: Filament-current feedback out of range.
Possible Causes
Filament-current digital-to-analog/feedback value was outside expected range.
Troubleshooting Steps
Check mAs/filament calibration, feedback circuitry and tube connections.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error
For Related Models
AMX-4, AMX-4+, Definium AMX 700
Error Number
CAL TAP ERROR 6
Error Code System Message
CAL TAP ERROR 6 - Filament-current feedback out of range
Question
What does CAL TAP ERROR 6 mean on a GE AMX-Series mobile X-ray?
Short Answer
CAL TAP ERROR 6: Filament-current feedback out of range.
Possible Causes
Filament-current feedback was outside the allowed calibration window.
Troubleshooting Steps
Check filament feedback, mAs calibration, tube/cathode connections and repeat tap calibration.
Parts Areas to Check
CPU board; calibration data; subsystem identified by the error