AMX Troubleshooting & Technical Support
Results Count
Error Number
Error 0x0909
Error Code System Message
Question
What does 0x0909 mean on a GE Brivo XR285amx?
Short Answer
0x0909: Rotation firmware not compatible with application firmware.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: Rotation firmware not compatible with application firmware.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x090A
Error Code System Message
Question
What does 0x090A mean on a GE Brivo XR285amx?
Short Answer
0x090A: Boot corrupted.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: Boot corrupted.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x090B
Error Code System Message
Question
What does 0x090B mean on a GE Brivo XR285amx?
Short Answer
0x090B: RTL2 is rising at bootup.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: RTL2 is rising at bootup.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x090C
Error Code System Message
Question
What does 0x090C mean on a GE Brivo XR285amx?
Short Answer
0x090C: ST10 watchdog error.
Possible Causes
The Djinn Beetle generator reported a Boot SW fault: ST10 watchdog error.
Troubleshooting Steps
Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.
Parts Areas to Check
For Related Models
Brivo XR285amx
Error Number
Error 0x0A81
Error Code System Message
Question
What does 0x0A81 mean on a GE Brivo XR285amx?
Short Answer
0x0A81: Internal power supplies out of specification.
Possible Causes
The Djinn Beetle generator reported a Various SW fault: Internal power supplies out of specification.
Troubleshooting Steps
Run System Sanity and Power Supplies, then review the error log. Check the indicated internal power, NVRAM/RTC, FPGA communication, or reboot condition before isolating an FRU.
Parts Areas to Check