top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 0x0909

Error Code System Message

​

0x0909, 0X0909, Brivo XR285amx 0x0909, Djinn Boot error, Rotation firmware not compatible with application firmware
Question

What does 0x0909 mean on a GE Brivo XR285amx?

Short Answer

0x0909: Rotation firmware not compatible with application firmware.

Possible Causes

The Djinn Beetle generator reported a Boot SW fault: Rotation firmware not compatible with application firmware.

Troubleshooting Steps

Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x090A

Error Code System Message

​

0x090A, 0X090A, Brivo XR285amx 0x090A, Djinn Boot error, Boot corrupted
Question

What does 0x090A mean on a GE Brivo XR285amx?

Short Answer

0x090A: Boot corrupted.

Possible Causes

The Djinn Beetle generator reported a Boot SW fault: Boot corrupted.

Troubleshooting Steps

Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x090B

Error Code System Message

​

0x090B, 0X090B, Brivo XR285amx 0x090B, Djinn Boot error, RTL2 is rising at bootup
Question

What does 0x090B mean on a GE Brivo XR285amx?

Short Answer

0x090B: RTL2 is rising at bootup.

Possible Causes

The Djinn Beetle generator reported a Boot SW fault: RTL2 is rising at bootup.

Troubleshooting Steps

Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x090C

Error Code System Message

​

0x090C, 0X090C, Brivo XR285amx 0x090C, Djinn Boot error, ST10 watchdog error
Question

What does 0x090C mean on a GE Brivo XR285amx?

Short Answer

0x090C: ST10 watchdog error.

Possible Causes

The Djinn Beetle generator reported a Boot SW fault: ST10 watchdog error.

Troubleshooting Steps

Review BDM/system self-recovery status and verify the Djinn software, database, FPGA, and rotation firmware are compatible. Load the correct 15 kW or 30 kW generator software/database when required.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Error 0x0A81

Error Code System Message

​

0x0A81, 0X0A81, Brivo XR285amx 0x0A81, Djinn Various error, Internal power supplies out of specification
Question

What does 0x0A81 mean on a GE Brivo XR285amx?

Short Answer

0x0A81: Internal power supplies out of specification.

Possible Causes

The Djinn Beetle generator reported a Various SW fault: Internal power supplies out of specification.

Troubleshooting Steps

Run System Sanity and Power Supplies, then review the error log. Check the indicated internal power, NVRAM/RTC, FPGA communication, or reboot condition before isolating an FRU.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page