top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

No generator power

Error Code System Message

​

Brivo XR285 generator 48V, LVLE2, power supplies
Question

What does the Djinn Power Supplies diagnostic check?

Short Answer

It checks the internal ±13 V, 48 V, and 3.3 V generator supplies while modules are off.

Possible Causes

Generator internal supply fault; 48 V LVLE2 issue.

Troubleshooting Steps

Run Power Supplies. For 48 V faults, verify LVLE2 output is 48 VDC ±2.5 V with the specified test setup and inspect the CAN/power connections.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Power Supply overheat or too high

Error Code System Message

​

Brivo XR285 thermal error, power supply temperature
Question

What does the Power Supply Thermal test check on the Brivo XR285amx?

Short Answer

It compares internal power-supply temperatures against allowed limits.

Possible Causes

Overtemperature; failed fan/cooling; temperature sensor issue.

Troubleshooting Steps

Run the Power Supply Thermal diagnostic. If temperature errors are present, check cooling, fans, sensors, and the applicable Djinn thermal error code before replacing a module.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Power control comm issues

Error Code System Message

​

Brivo XR285 Power Control Comm, Spyder communication
Question

What does the Brivo XR285amx Power Control Comm diagnostic test?

Short Answer

It checks low-level communication and control between the Spyder control board and other hardware controllers.

Possible Causes

Spyder or controller communication fault; cable or board issue.

Troubleshooting Steps

Run Power Control Comm in SUIF. If it fails, use the error log and System Sanity/communication results to isolate the controller, wiring, or board that is not communicating.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

System sanity issues

Error Code System Message

​

Brivo XR285 System Sanity, Spyder CRC memory test
Question

What does System Sanity test on the Brivo XR285amx?

Short Answer

System Sanity checks board voltages, Spyder memory, flash-storage CRC integrity, and local/system communications.

Possible Causes

Board supply voltage issue; Spyder memory fault; corrupted software; communication fault.

Troubleshooting Steps

Run System Sanity and identify which subtest fails. Use the result to isolate voltage, memory, software-integrity, or communication problems before replacing hardware.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

Error Number

Tube park interlock not working

Error Code System Message

​

Brivo XR285 tube park interlock
Question

How do I test the tube park interlock on the Brivo XR285amx?

Short Answer

Tube Interlocks verifies that the system recognizes the tube park as latched.

Possible Causes

Tube-park latch/sensor fault; wiring or system-input problem.

Troubleshooting Steps

Run Tube Interlocks. If the latch state is not recognized, inspect the tube-park latch, sensor, wiring, and system-control input.

Parts Areas to Check

​

For Related Models

Brivo XR285amx

bottom of page