AMX Troubleshooting & Technical Support

Results Count

Error Number

Calibration checksum blocks application mode

Error Code System Message

TEST 04 FAILED / CHECK SUM ERR x

AMX calibration checksum, TEST 04 FAILED
Question

Can a calibration checksum failure prevent normal AMX-Series operation?

Short Answer

Yes. Calibration checksum errors can prevent entry into normal application mode until the calibration data or affected calibration is corrected.

Possible Causes

Corrupted or invalid calibration data; interrupted calibration.

Troubleshooting Steps

Identify the failing checksum. Enter service/calibration mode only as directed, repair the affected calibration, and rerun the relevant checksum/test before normal use.

Parts Areas to Check

Calibration data; CPU non-volatile memory

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Cannot Enter “SERVICE MODE” - But Portable Is Operational In Applications Mode

Error Code System Message

​

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0170, Cannot Enter “SERVICE MODE” - But Portable Is Operational In Applications Mode
Question

How do I troubleshoot this AMX-4 Series issue: Cannot Enter “SERVICE MODE” - But Portable Is Operational In Applications Mode?

Short Answer

Bad Display Controller Board

Possible Causes

Problem: Cannot Enter “SERVICE MODE” - But Portable Is Operational In Applications Mode

Troubleshooting Steps

Bad Display Controller Board

Parts Areas to Check

Display Controller Board

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Cannot calibrate kVp (gets hung), filament taps or filament current table (low filament error) etc., or load data in data base access properly, can’t enter calibration mode or getting false error codes displayed, after c

Error Code System Message

​

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0149, Cannot calibrate kVp (gets hung), filament taps or filament current table (low filament error) etc., or load data in data base access properly, can’t enter calibration mode or getting false error codes displayed, after c
Question

How do I troubleshoot this AMX-4 Series issue: Cannot calibrate kVp (gets hung), filament taps or filament current table (low filament error) etc., or load data in data base access properly, can’t enter calibration mode or getting false error codes displayed, after c?

Short Answer

Remove the two NOVRAM chips (U54 & U107) on the newer CPU Board and replace them with the two NOVRAM chips from the older CPU Board, if it is the same style of board. If the board is the “older” style of CPU Board that has only one NOVRAM chip, Then re-order a new CPU Board (Orderable Part # 46-264974G5 or 46-264974G6), or replace the NOVRAM chip(s) (Orderable Part # 46-230318P1)

Possible Causes

Problem: Cannot calibrate kVp (gets hung), filament taps or filament current table (low filament error) etc., or load data in data base access properly, can’t enter calibration mode or getting false error codes displayed, after changing out the CPU Board (part # T3497H, 46-264974G2 or later)

Troubleshooting Steps

Remove the two NOVRAM chips (U54 & U107) on the newer CPU Board and replace them with the two NOVRAM chips from the older CPU Board, if it is the same style of board. If the board is the “older” style of CPU Board that has only one NOVRAM chip, Then re-order a new CPU Board (Orderable Part # 46-264974G5 or 46-264974G6), or replace the NOVRAM chip(s) (Orderable Part # 46-230318P1)

Parts Areas to Check

CPU Board

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Capacity bar graph disagrees with battery voltage

Error Code System Message

CHARGE COMPLETE

AMX capacity meter wrong, bar graph wrong, CHARGE COMPLETE
Question

How do I synchronize the AMX-Series battery capacity meter with the displayed capacity?

Short Answer

The manual provides two methods: complete a full charge to CHARGE COMPLETE, or use the service LOOP TEST procedure to copy internal capacity data to the displayed capacity.

Possible Causes

Capacity-display drift after charging or battery replacement; stale internal capacity data.

Troubleshooting Steps

Prefer a full charge through CHARGE COMPLETE. If service synchronization is required, use the manual's service procedure to synchronize internal and displayed capacity.

Parts Areas to Check

Capacity meter; CPU non-volatile memory; batteries

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Charger calibration required

Error Code System Message

CAL CHGR ERR / CHARG BYTE/WORD LIMIT

AMX charger calibration, CAL CHGR ERR
Question

What does AMX-Series battery-charger calibration learn?

Short Answer

The charger calibration establishes the relationship between commanded charge current and measured charge-current feedback so actual charging current can be interpreted correctly.

Possible Causes

Charger-current feedback/calibration mismatch.

Troubleshooting Steps

Perform charger calibration when CHARG/CAL CHGR errors occur or after relevant charger/CPU service, then verify current feedback across the required ranges.

Parts Areas to Check

Charger board; CPU board; current feedback

For Related Models

AMX-4, AMX-4+, Definium AMX 700