AMX Troubleshooting & Technical Support

Results Count

Error Number

Detector unparked causes slow drive

Error Code System Message

Exposure inhibit

AMX 700 slow drive detector unparked, exposure inhibit
Question

Why does the Definium AMX 700 drive slowly when the detector is unparked?

Short Answer

The functional test specifies slow drive with the detector unparked and an exposure inhibit during an exam.

Possible Causes

Normal safety/interlock behavior when detector is not parked.

Troubleshooting Steps

Park the detector for normal driving. During service testing, verify slow drive when unparked and verify an exposure inhibit appears during an exam.

Parts Areas to Check

Detector park switch; Beetle Board; drive/interlock logic

For Related Models

Definium AMX 700

Error Number

Digital Mode - PC protocol communication established

Error Code System Message

Digital Mode - PC protocol communication established

Digital Mode, AMX 700, PC protocol communication established
Question

What does “Digital Mode” mean on a GE Definium AMX 700?

Short Answer

Digital Mode: PC protocol communication established.

Possible Causes

Subsystem/category: PC / Beetle Communication. Source classification: STATUS.

Troubleshooting Steps

If Digital Mode does not appear after PC boot, check PC boot, Beetle heartbeats and interface cables.

Parts Areas to Check

Mobile PC; Beetle Board

For Related Models

Definium AMX 700

Error Number

Digital power supply does not function

Error Code System Message

​

AMX 700 digital power supply fuse orientation, no power
Question

Why can the Definium AMX 700 Digital Detector Power Supply fail after fuse replacement?

Short Answer

The manual states the fuses must be installed in the correct orientation or the power supply may not function.

Possible Causes

Fuse installed in the wrong orientation; blown fuse; power-supply fault.

Troubleshooting Steps

Verify each Digital Detector Power Supply fuse is installed in the orientation shown in the manual. If power is still absent, continue with debug-header voltage and fuse checks.

Parts Areas to Check

Digital Detector Power Supply; 2A/3A/5A fuses

For Related Models

Definium AMX 700

Error Number

Direct laptop ping works but hospital network does not

Error Code System Message

​

AMX 700 direct ping works hospital network fails
Question

What does it mean if the Definium AMX 700 can ping a service laptop but not the hospital network?

Short Answer

The manual states this points toward incorrect logical network information or a hospital-network infrastructure issue.

Possible Causes

Hospital network configuration/infrastructure issue rather than local Ethernet hardware.

Troubleshooting Steps

Re-verify system IP, subnet and router settings, then have the site contact/network administrator verify the hospital network infrastructure.

Parts Areas to Check

System NIC; hospital network

For Related Models

Definium AMX 700

Error Number

Drive handle does not enable drive

Error Code System Message

​

AMX 700 will not drive, drive handle no response
Question

What should I check if the Definium AMX 700 does not drive when I squeeze the handle?

Short Answer

The Beetle Board functional test expects drive enable when the drive handle is squeezed.

Possible Causes

Drive-enable signal not recognized; Beetle Interface Board or AMX drive-system issue.

Troubleshooting Steps

Verify the Beetle Board is operating normally, then squeeze the drive handle and confirm drive enable. If it fails, isolate the interface/drive-enable path before replacing parts.

Parts Areas to Check

Drive handle; Beetle Board; AMX drive system

For Related Models

Definium AMX 700