  • GE 46-208701P4 Clamp Wraparound Mounting Bracket | AMCI Medical…

    GE 46-208701P4 Clamp Wraparound Mounting Bracket for AMX-4+, Definium AMX 700. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 46-208701P4 Part Type Metal Section Area Generator Mid Section Part Condition New, Harvested Compatible Model AMX-4+, Definium AMX 700 Part Notes Previous Next Request price quote < Search more 46-208701P4 Clamp Wraparound Mounting Bracket Part Description

  • Error 45F | AMCI Medical Imaging

    GE mobile X-ray troubleshooting for Error 45F. Review possible causes, diagnostic steps and service guidance from AMCI Medical Imaging. Troubleshooting AMX Error Number Error 45F Error Code / System Message ERROR 45F - Filament-short signal remains active after filaments removed Error Question What does ERROR 45F mean on a GE AMX-Series mobile X-ray? Short Answer ERROR 45F: Filament-short signal remains active after filaments removed. Possible Causes Short-detect remained active after filament loads were removed. Troubleshooting Steps Inspect high-voltage cable polarity/orientation and filament short-detect circuitry. Parts Areas to Check Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault < Search more Previous Next For Related Models AMX-4, AMX-4+, Definium AMX 700

  • FlashPad detector charging stop during image acquisition | AMCI Medical Imaging

    GE mobile X-ray troubleshooting for FlashPad detector charging stop during image acquisition. Review possible causes, diagnostic steps and service guidance from AMCI Medical Imaging. Troubleshooting AMX Error Number FlashPad detector charging stop during image acquisition Error Code / System Message Error Question Why does detector charging stop during image acquisition? Short Answer The manual states detector-battery charging stops during image acquisition and resumes in standby/charging conditions. Possible Causes Normal detector behavior. Troubleshooting Steps Allow the detector to return to standby/charging mode. If charging does not resume afterward, test the charger/bin/docking path and battery condition. Parts Areas to Check < Search more Previous Next For Related Models Optima XR200amx, Optima XR220amx

  • GE 5832964 Thorax Column Bulkhead - Collapsible Column | AMCI…

    GE 5832964 Thorax Column Bulkhead - Collapsible Column for AMX Navigate. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 5832964 Part Type Assembly Section Area Thorax Mid Section Part Condition Harvested, Refurbished Compatible Model AMX Navigate Part Notes Previous Next Request price quote < Search more 5832964 Thorax Column Bulkhead - Collapsible Column Part Description

  • MPPS updates not sent | AMCI Medical Imaging

    GE mobile X-ray troubleshooting for MPPS updates not sent. Review possible causes, diagnostic steps and service guidance from AMCI Medical Imaging. Troubleshooting AMX Error Number MPPS updates not sent Error Code / System Message Error Question Why can’t the Definium AMX 700 send MPPS updates? Short Answer Verify networking and the MPPS server IP address, case-sensitive AE Title and DICOM port, then run C-Echo. Possible Causes Network issue; MPPS configuration mismatch; server unavailable. Troubleshooting Steps Open Network Connections, select the MPPS device under HIS/RIS, verify IP/AE Title/port, run C-Echo and review the system error log if the test fails. Parts Areas to Check MPPS server; AMX 700 DICOM configuration < Search more Previous Next For Related Models Definium AMX 700

  • GE C2001-M3-80 Flat Washer M3 Black Oxide | AMCI Medical Imaging

    GE C2001-M3-80 Flat Washer M3 Black Oxide for Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate. Available from AMCI Medical… GE Healthcare Part No. C2001-M3-80 Part Type Fastener Retainer Section Area Collimator Part Condition New Compatible Model Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate Part Notes Previous Next Request price quote < Search more C2001-M3-80 Flat Washer M3 Black Oxide Part Description

  • Error 0x060F | AMCI Medical Imaging

    GE mobile X-ray troubleshooting for Error 0x060F. Review possible causes, diagnostic steps and service guidance from AMCI Medical Imaging. Troubleshooting AMX Error Number Error 0x060F Error Code / System Message Error Question What does 0x060F mean on a GE Brivo XR285amx? Short Answer 0x060F: Watchdog firmware error. Possible Causes The Djinn Beetle generator reported a kV FW fault: Watchdog firmware error. Troubleshooting Steps Run KV Measurement, KV Test, Inverter Test, and Resonant Components as applicable. Check the HV Tank-to-Power Module measurement cable, DC-bus error data, inverter/tank connections, and HV cables for arcing when indicated. Parts Areas to Check < Search more Previous Next For Related Models Brivo XR285amx

  • Error 12 | AMCI Medical Imaging

    GE mobile X-ray troubleshooting for Error 12. Review possible causes, diagnostic steps and service guidance from AMCI Medical Imaging. Troubleshooting AMX Error Number Error 12 Error Code / System Message ERROR 12 - Left drive stall / release-handle fault Error Question What does ERROR 12 mean on a GE AMX-Series mobile X-ray? Short Answer ERROR 12: Left drive stall / release-handle fault. Possible Causes Left stall signal detected. Troubleshooting Steps Check motor relays/snubber circuits, brakes, drive motor and stall/release wiring. Parts Areas to Check Drive control board; drive handle/Hall sensors; motor relays; drive motors; feedback wiring < Search more Previous Next For Related Models AMX-4, AMX-4+, Definium AMX 700

  • Error 437 | AMCI Medical Imaging

    GE mobile X-ray troubleshooting for Error 437. Review possible causes, diagnostic steps and service guidance from AMCI Medical Imaging. Troubleshooting AMX Error Number Error 437 Error Code / System Message ERROR 437 - Back-up timer OK missing during pre-exposure check Error Question What does ERROR 437 mean on a GE AMX-Series mobile X-ray? Short Answer ERROR 437: Back-up timer OK missing during pre-exposure check. Possible Causes Backup-timer OK status was not present during prep. Troubleshooting Steps Check backup timer/status signal and CPU critical-status input. Parts Areas to Check Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault < Search more Previous Next For Related Models AMX-4, AMX-4+, Definium AMX 700

  • GE 46-154083P2 Resistor 200W 20 | AMCI Medical Imaging

    GE 46-154083P2 Resistor 200W 20 for AMX-4, AMX-4+, Definium AMX 700. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 46-154083P2 Part Type Electronic Component Section Area Generator Mid Section Part Condition New Compatible Model AMX-4, AMX-4+, Definium AMX 700 Part Notes Previous Next Request price quote < Search more 46-154083P2 Resistor 200W 20.5 Ohms 10% Part Description

  • GE 5557037 Cable Assembly Mantis to Filter | AMCI Medical Imaging

    GE 5557037 Cable Assembly Mantis to Filter for Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate. Available from AMCI Medical… GE Healthcare Part No. 5557037 Part Type Cable Assembly Section Area Thorax Mid Section Part Condition Harvested Compatible Model Brivo XR285amx, Optima XR200amx, Optima XR220amx, Optima XR240amx, AMX Navigate Part Notes Previous Next Request price quote < Search more 5557037 Cable Assembly Mantis to Filter Part Description

  • GE 514A146P1 Fuse 10A 250V 3AB | AMCI Medical Imaging

    GE 514A146P1 Fuse 10A 250V 3AB for AMX-4. Available from AMCI Medical Imaging for mobile X-ray service and replacement. GE Healthcare Part No. 514A146P1 Part Type Electronic Component Section Area Generator Mid Section Part Condition New, Harvested Compatible Model AMX-4 Part Notes Previous Next Request price quote < Search more 514A146P1 Fuse 10A 250V 3AB Part Description