top of page

AMX Troubleshooting & Technical Support

Results Count

Error Number

Error 433

Error Code System Message

ERROR 433 - Leakage-compensation command feedback high

ERROR 433, Leakage-compensation command feedback high, AMX-4 ERROR 433, AMX-4+ ERROR 433, Definium AMX 700 ERROR 433, GE AMX error ERROR 433
Question

What does ERROR 433 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 433: Leakage-compensation command feedback high.

Possible Causes

Leakage-compensation feedback was too high during prep.

Troubleshooting Steps

Check leakage compensation feedback, filament board and kVp feedback.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 435

Error Code System Message

ERROR 435 - Tap feedback does not match selected tap

ERROR 435, Tap feedback does not match selected tap, AMX-4 ERROR 435, AMX-4+ ERROR 435, Definium AMX 700 ERROR 435, GE AMX error ERROR 435
Question

What does ERROR 435 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 435: Tap feedback does not match selected tap.

Possible Causes

Selected tap feedback was inconsistent during prep.

Troubleshooting Steps

Check tap-frequency/feedback signals, tap relay board and generator control.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 435

Error Code System Message

435

AMX-4, AMX-4+, Definium AMX 700, ERROR 435, Error 435
Question

What should I check for error 435 on the AMX-4 Series?

Short Answer

1) Check that all contactors K1 thru K6 an AMX1 A4 panel pull in when PREP is pressed 2) Check installation of ribbon cable from 1 kHz Inverter Board to CPU Board or opto-couplers on 1 kHz Inverter Board

Possible Causes

1) Check that all contactors K1 thru K6 tap relays pull in when PREP is pressed (look closely at K4, K5, K6 relays) 2) Check installation of ribbon cable from 1 kHz Inverter Board to CPU Board or opto-couplers on 1 kHz Inverter Board 3) Bad CPU and 1kHz Broad

Troubleshooting Steps

1) Check that all contactors K1 thru K6 an AMX1 A4 panel pull in when PREP is pressed 2) Check installation of ribbon cable from 1 kHz Inverter Board to CPU Board or opto-couplers on 1 kHz Inverter Board

Parts Areas to Check

CPU Board; 1 kHz Inverter Board

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 436

Error Code System Message

436

AMX-4, AMX-4+, Definium AMX 700, ERROR 436, Error 436
Question

What should I check for error 436 on the AMX-4 Series?

Short Answer

1) Defective Filament kVp Control Board

Possible Causes

1) Defective Filament kVp Control Board

Troubleshooting Steps

1) Defective Filament kVp Control Board

Parts Areas to Check

​

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Error 436

Error Code System Message

ERROR 436 - X-RAY ON asserted during pre-exposure interlock check

ERROR 436, X-RAY ON asserted during pre-exposure interlock check, AMX-4 ERROR 436, AMX-4+ ERROR 436, Definium AMX 700 ERROR 436, GE AMX error ERROR 436
Question

What does ERROR 436 mean on a GE AMX-Series mobile X-ray?

Short Answer

ERROR 436: X-RAY ON asserted during pre-exposure interlock check.

Possible Causes

X-RAY ON was active before exposure should begin.

Troubleshooting Steps

Check X-RAY ON status line, CPU critical-status input and generator-control sequencing.

Parts Areas to Check

Generator control board; CPU board; inverter/rotor/filament/tap/HV subsystem indicated by the fault

For Related Models

AMX-4, AMX-4+, Definium AMX 700

bottom of page