AMX Troubleshooting & Technical Support

Results Count

Error Number

Need to enter service mode

Error Code System Message

​

AMX service mode, calibration mode, extended diagnostics
Question

What is AMX-Series service mode used for?

Short Answer

Service mode is intended for service personnel and provides access to calibration, extended diagnostics, data logs, and data-base access.

Possible Causes

Service/maintenance task.

Troubleshooting Steps

Use the service switch and approved service procedure. Restrict calibration and data-base changes to qualified personnel because incorrect values can affect system operation and exposure output.

Parts Areas to Check

CPU; calibration database; diagnostic tools

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

No Display After Installing ESD Kit

Error Code System Message

​

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0158, No Display After Installing ESD Kit
Question

How do I troubleshoot this AMX-4 Series issue: No Display After Installing ESD Kit?

Short Answer

Board Mounting Screw(s) Retaining Nut Lock-washers Impinge On The +40 VDC Bus OnThe Display Controller Board (46-232832G1). The +40 VDC Bus Is Too Close To The Retaining Nut Lock-washers—Shorts Out The +40 VDC Bus When The ESD Kit Ground Cables Are Added. Reposition Existing Mounting Screws/Lock-washers. CQA 972174

Possible Causes

10) Problem: No Display After Installing ESD Kit

Troubleshooting Steps

Board Mounting Screw(s) Retaining Nut Lock-washers Impinge On The +40 VDC Bus OnThe Display Controller Board (46-232832G1). The +40 VDC Bus Is Too Close To The Retaining Nut Lock-washers—Shorts Out The +40 VDC Bus When The ESD Kit Ground Cables Are Added. Reposition Existing Mounting Screws/Lock-washers. CQA 972174

Parts Areas to Check

Display Controller Board

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

No “normal” kVp waveform can be viewed (Applications Or Check-Sum Bypass) With Errors 439 Or 469

Error Code System Message

​

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0164, No “normal” kVp waveform can be viewed (Applications Or Check-Sum Bypass) With Errors 439 Or 469
Question

How do I troubleshoot this AMX-4 Series issue: No “normal” kVp waveform can be viewed (Applications Or Check-Sum Bypass) With Errors 439 Or 469?

Short Answer

Temporarily Jumper TP-14 To TP-15 On Filament kVp Board To Obtain A “normal” kVp Waveform

Possible Causes

16) Problem: No “normal” kVp waveform can be viewed (Applications Or Check-Sum Bypass) With Errors 439 Or 469

Troubleshooting Steps

Temporarily Jumper TP-14 To TP-15 On Filament kVp Board To Obtain A “normal” kVp Waveform

Parts Areas to Check

Filament kVp Board

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Open filament in X-Ray Tube (Error 45F) during PREP. Defective Filament kVp Control Board (Q170, Q160 or Q173 shorted, Inverter running at maximum output. CQA 964232.

Error Code System Message

45F

AMX-4, AMX-4+, Definium AMX 700, AMCI-AMX4-TS-0151, Open filament in X-Ray Tube (Error 45F) during PREP. Defective Filament kVp Control Board (Q170, Q160 or Q173 shorted, Inverter running at maximum output. CQA 964232.
Question

How do I troubleshoot this AMX-4 Series issue: Open filament in X-Ray Tube (Error 45F) during PREP. Defective Filament kVp Control Board (Q170, Q160 or Q173 shorted, Inverter running at maximum output. CQA 964232?

Short Answer

Replace Filament kVp Board

Possible Causes

3) Problem: Open filament in X-Ray Tube (Error 45F) during PREP. Defective Filament kVp Control Board (Q170, Q160 or Q173 shorted, Inverter running at maximum output. CQA 964232.

Troubleshooting Steps

Replace Filament kVp Board

Parts Areas to Check

Filament kVp Board; X-Ray Tube

For Related Models

AMX-4, AMX-4+, Definium AMX 700

Error Number

Oscilloscope triggering during calibration

Error Code System Message

​

AMX oscilloscope trigger calibration, CALIBRATE KVP
Question

How should I trigger an oscilloscope for AMX-Series generator calibration?

Short Answer

The manual recommends establishing the scope trigger before entering CALIBRATE KVP because the unit fails power-up tests in calibration mode.

Possible Causes

Difficult oscilloscope triggering during service calibration.

Troubleshooting Steps

Set up the trigger in normal operating mode using the specified low-kVp/mAs test exposure, then enter calibration with the scope already triggering reliably.

Parts Areas to Check

Generator calibration test points

For Related Models

AMX-4, AMX-4+, Definium AMX 700